EasyManua.ls Logo

Fluke PM6690 - Profiling

Fluke PM6690
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
a duration that expires in the gap between
pulse #3 and #4.
Use the same test setup as in the preceding ex
-
ample. Then proceed as follows:
–
Press SETTINGS ® Arm and select
Stop Chan E and negative Stop Slope.
–
Measure.
#4 Profiling
Profiling means measuring frequency versus
time. Examples are measuring warm-up drift
in signal sources over hours, measuring the
linearity of a frequency sweep during seconds,
VCO switching characteristics during milli
-
seconds, or the frequency changes inside a
“chirp radar” pulse during microseconds.
These counters can handle many profiling
measurement situations with some limitations.
Profiling can theoretically be done manually,
i.e., by reading individual measurement re
-
sults and plotting in a graph. However, to
avoid getting bored long before reaching your
800th or so measurement result, you must use
some computing power and a bus interface. In
profiling applications, the counter acts as a
fast, high-resolution sampling front end, stor
-
ing results in its internal memory. These re
-
sults are later transferred to the controller for
analysis and graphical presentation. The
TimeView™ software package greatly simpli
-
fies profiling.
You must distinguish between two different
types of measurements called free-running
and repetitive sampling.
n
Free-Running Measurements
Free-running measurements are performed
over a longer period, e.g., to measure the sta
-
bility over 24 hours of oscillators, to measure
initial drift of a generator during a 30-minute
warm-up time, or to measure short-term sta
-
bility during 1 or 10s. In these cases, measure
-
ments are performed at user-selected intervals
in the range 2 ms to 1000 s. There are several
different ways of performing the measure
-
ments at regular intervals.
Measurements using the statistics
features for setting the “pacing time”
By setting the pacing time to 10 s for exam-
ple, measurements are automatically made at
10 s intervals until the set number of samples
has been taken. The range is 2 - 2
*
10
9
.Use
HOLD/RUN and RESTART if you want to
stop after one full cycle. You can watch the
trend or spread on the graphic display while
the measurement is proceeding.
Using a controller as a “pacer”
As an alternative, the timer in the controller
can be used for pacing the individual measure
-
ments. This allows for synchronization with
external events, for instance a change of DUT
when checking a series of components.
Using external arming signals
External arming signals can also be used for
“pacing.” For example with an arming signal
consisting of 10 Hz pulses, individual mea
-
surements are armed at 100 ms intervals.
Letting the counter run free
When the counter is free-running, the shortest
delay between measurements is approximately
4 ms (internal calibration OFF) or 8 ms (inter
-
Arming Examples 5-13
Measurement Control
S y n c E
S t a r t A r m i n g
I n p u t A
S t o p A r m i n g
Fig. 5-15 Using both start and stop arm
-
ing to select the part of the
burst that is of interest.

Table of Contents