EasyManua.ls Logo

Fluke PM6690 - Frequency Reference Calibration

Fluke PM6690
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Phase
LSD
displ
= 0.001° to 1°, depending on input
frequency (see Phase measurement spec.)
Ratio f
1
/f
2
LSD
displ
= Pr escaler factor / (f
2
x meas. time)
Volt
LSD
displ
=TBD
Number of Pulses in Burst
LSD
displ
=1
Time Interval, Pulse Width,
Rise/Fall Time
n
Total Random Uncertainty rms
uE
q
StartTrigger Error StopTrigger Error=+ +
222
()()
[s] or minimum 1 ps
n
Total Systematic Uncertainty
u = ± Trigger Level Timing Error
± 500 ps Systematic Error
±TimeBaseErrorxMeasurementResult
[s]
Frequency & Period
n
Total Random Uncertainty rms
For measuring times <200 ms:
u
E
q
Start Trigger E rror
Measuring Time
=
+
´
2
2
2
*( )
´ Measurement Result
[Hz or s]
For measuring times >=200 ms:
u
E
q
Start Trigger E rror
Measuring Time N
=
+
*
´
2
2
2
2
*( )
´ Measurement Result
[Hz or s]
N = 800 /Measuring Time, however always
>=6 and <=1000
n
Total Systematic Uncertainty
u=±TimeBaseErrorxMeasurementResult±
200 ps
/Measuring Time [Hz or s]
Ratio f
1
/f
2
n
Total Random Uncertainty rms
u
Presc Fact f Strt Trg Err of f
f
=
+´ ´(..) ( )
2
2
12
2
2
[dimensionless, e.g. ppm]
Phase
n
Total Random Uncertainty rms
uE
q
Strt Trg Err Stop Trg Err Freq=+ + ´´
´°
22 2
360
()( ).
[]
or minimum: (1 ps) x Frequency x 360 [°]
n
Total Systematic Uncertainty
u = ± Trigger Level Timing Error
± 500 ps Systematic Error x Freq. x 360 [°]
Duty Factor
n
Total Random Uncertainty rms
uE
q
Strt Trg Err Stop Trg Err Freq=+ + ´
22 2
()( )
.
or minimum: (1 ps) x Frequency
[dimensionless, e.g ppm]
n
Total Systematic Uncertainty
u = ± Trigger Level Timing Error x Frequency
± 500 ps Systematic Error x Frequency
[dimensionless, e.g. ppm]
Specifications
8-10 Measurement Uncertainties

Table of Contents

Other manuals for Fluke PM6690

Related product manuals