EasyManua.ls Logo

Fluke PM6690 - Explanations; Timebase Options

Fluke PM6690
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
8-12 Timebase Options
Product Family '90'
Option Standard PM6690/_5_ PM6690/_6_
Timebase Type UCXO OCXO OCXO
Uncertainty due to:
-Calibration adjustment tolerance
@ +23 °C ± 3°C
<1x10
-6
<1x10
-8
<3x10
-9
-Aging
per 24 h
per month
per year
1)
<5x10
-7
<5x10
-6
2)
<5x10
-10
<1x10
-8
<5x10
-8
2)
<3x10
-10
<3x10
-9
<1.5x10
-8
-Temperature variation:
0 °C - 50 °C
20 °C - 26 °C (typ. values)
<1x10
-5
<3x10
-6
<5x10
-9
<1x10
-9
<2.5x10
-9
<4x10
-10
-Power voltage variation: ±10%
<1x10
-8
<5x10
-10
<5x10
-10
Short term stability:
(Root Allan Variance)
t =1s
Typical values t =10s
not specified 1x10
-11
1x10
-11
5x10
-12
5x10
-12
Power-on stability:
-Deviation versus final value after 24 h
on time, after a warm-up time of
:
1)
30 min
<1x10
-8
10 min
<5x10
-9
10 min
Total uncertainty,
for operating
temperature 0°C to 50 °C,
@2s (95 %) confidence interval:
-1 year after calibration
-2 years after calibration
<1.2x10
-5
<1.5x10
-5
<6x10
-8
<1.2x10
-7
<1.8x10
-8
<3.6x10
-8
Typical total uncertainty, for op
-
erating temperature 20°C to 26 °C
,
@2s (95 %) confidence interval:
-1 year after calibration
-2 years after calibration
<7x10
-6
<1.2x10
-5
<6x10
-8
<1.2x10
-7
<1.7x10
-8
<3.5x10
-8
Explanations
1)
Negligible in comparison with the deviation caused by 1 °C temperature change.
2)
After 1 month of continuous operation.
NOTE: Electrical adjustment by means of tuning voltage from DAC; no potentiometer trimming.
Serial interface to all optional oscillators for closed-case calibration and status reporting.
UCXO: Uncompensated Crystal Oscillator
OCXO: Oven Controlled Crystal Oscillator
Timebase Options

Table of Contents

Other manuals for Fluke PM6690

Related product manuals