EasyManua.ls Logo

Flybox Vigilus - Page 48

Default Icon
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Flybox
VIGILUS - Operating manual
48
®
Rev. 1.0
Fuel computer menu
K factor: Set the fuel flow transducer's K-factor, if installed. The
K-factor of a fuel flow transducer is the number of electric pulses
for 1 gallon of fuel consumption. If you have the K-factor in liters,
multiply the value by 3.78 before entering it. Default value is
416400 (nominal k-factor of Flybox® TFTHP flow transducer).
NOTE: For the Rotax 912iS engine the fuel flow is read by the
ECU so this parameter isn’t used.
Filter: This parameter affect the readings and the gauges
displayed: a low value means that the readings will be more
fast and unfiltered (but subject to fluctuations), an high value
means that the readings will be more slow and stable (range:
0~999, default=100).
Calibration: Calibration of the K-factor of the fuel flow
transducer, if installed. Refer to chapter 1.4 “Fuel flow
transducer calibration”.
NOTE: It's recommended to execute the K factor calibration
as soon as possible to have the maximum accuracy in the
fuel flow measurements.
Remain. Alm = ON/OFF: Enable or disable the alarm on
remaining fuel quantity (default=ON).
Remain. Q.ty: Set the fuel quantity for the “Remain. Alm”. When
the remaining fuel of the Fuel Computer is below this setpoint
the alarm is activated (range: 0.1~99.9 lt, default=30.0 lt).
Remain. Rep: Set the number of times that the vocal alarm for
the remaining fuel quantity is repeated on the audio output
(range:1~5, default=1).
Instrument configuration