Q2ec-FM0049
Move the table.
Use the light localizer to determine the center
position for imaging. Next, hold down the
(SET) button on the gantry control panel to
move the table until the position set by the light
localizer is aligned with the isocenter.
5
1
Place the pads and coil on the table.
2
3
To prevent eye damage from the laser
beam, have patients close their eyes or
take other appropriate measures.
3
When you move the table by
pushing the button,
table speed can be slower by
pushing it again.
Position the patient on the table, and adjust the
positions of the patient and the coil by moving
the pad or table horizontally until the region to be
scanned is aligned with the isocenter.
Chapter 3 Preparing for Imaging
3.1 Preparing the patient
Prepare the patient.
NOTICE
Pad