EasyManuals Logo
Home>GE>Laboratory Equipment>AKTAprocess

GE AKTAprocess User Manual

GE AKTAprocess
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
Possible cause/corrective actionProblemComponent
1
Open a System Control module.
2
Select the System:Connect menu command.
or
Click the Connect to system toolbar icon.
Result: The System Connect dialog box opens.
3
Select the system you want to connect.
4
Click OK.
No connection between
the system and UNICORN
Computer
Check circuit breaker, both in system and for external
supply, as applicable.
Power failure during a runPower
Check if there is insufficient air pressure, caused for
instance by malfunctioning air supply equipment or a
leaking air hose.
a) Rectify the air supply problem.
b) Restart the run by pressing the Continue button.
Compressed air failure
during a run. The alarm
“No air supply to the sys-
tem” is displayed and the
system enters Pause
mode.
Compressed air
Perform a valve calibration. If a Valve error is trig-
gered, a valve calibration may restore the valve func-
tion. For instructions on how to perform a valve calibra-
tion, see Maintenance annually or as needed, on
page 131. If a valve calibration does not help, contact
your local GE Healthcare representative.
Alarm Valve error is dis-
played
Valves
If the Valve error alarm is accompanied by No air
supply to the system, the alarm is caused by a com-
pressed air failure (see above).
158 ÄKTAprocess Operating Instructions 29-0252-49 AA
7 Troubleshooting

Table of Contents

Other manuals for GE AKTAprocess

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTAprocess and is the answer not in the manual?

GE AKTAprocess Specifications

General IconGeneral
BrandGE
ModelAKTAprocess
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals