Possible cause/corrective actionProblemComponent
Emergency button has been pressed.
Air pressure too low. Check the pressure of air
supply. Adjust if necessary.
No inlet or outlet valve open. Check method and
valves.
Incorrect method. Check by entering
Pump:ManFlow > 1%.
If none of the above, contact GE Healthcare service
personnel.
Pump not workingPump
Check that the connected inlet is actually used.
Inlet containers are placed too low relative to the
pump. Check inlet containers.
No liquid is supplied to the pump. Check inlet con-
tainers.
Tubing from inlet container causes pressure or flow
loss. Reasons may be too long tubing, too small in-
ternal diameter, tube may have a narrow section
or is partly plugged.
Malfunctioning valve at container battery limit.
Little or no flow
Check that connected outlet is actually used.
Outlet containers are placed too high relative to
the pump.
Tubing to outlet container causes pressure or flow
loss. Reasons may be too long tubing, too small in-
ternal diameter, tube may have a narrow section
or may be partly plugged. Check also for non-
functioning valve(s).
Malfunctioning valve at container battery limit.
Too high outlet pressurePump
ÄKTAprocess Operating Instructions 29-0252-49 AA 159
7 Troubleshooting

Table of Contents

Other manuals for GE AKTAprocess

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTAprocess and is the answer not in the manual?

GE AKTAprocess Specifications

General IconGeneral
BrandGE
ModelAKTAprocess
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals