September 2004
2-32 Programming Site Data
Setting Up a Totalizer
Output
1. At Output B or D, select OFF or TTLZR and press [ENTER].
2. Do one of the following:
• If you selected OFF, the meter returns to the Global I/O window.
Go to Procedure Options on page 2-32.
• If you are using a 1-channel meter, go to Step 4.
• If you are using a 2-channel/path meter, go to Step 3.
3. At Channel, press [
X] to select the desired channel/path and press
[ENTER].
4. Press [
X] to select the desired measurement parameter and press
[ENTER].
5. Enter a value for the minimum pulse on time (between 50 µsec
and 500,000 µsec) for the frequency of the totalizer pulses and
press [ENTER]. Press [
W] and [X] to move the cursor to desired
location and press
[S] and [T] to increase or decrease the
number.
Note: A complete pulse consists of equal amounts of ON and OFF
times. Choose a value that is compatible with the counter to be
used.
6. At UNITS/PULSE enter a value for the number of measurement
units represented by each pulse and press [ENTER].
Procedure Options After completing the above steps, the user program returns to the
Global I/O window. Do one of the following:
• To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
• To leave the user program, press [ESC] four times.