5-40 170 Series Monitor Revision C
2000947-004
Theory of Operation: FECG/IUP Board Theory of Operation
Figure 5-22. FECG/IUP Board Block Diagram
FECG Low Noise
Differential
Amplifier with
Protection Circuitry
RL Drive Circuitry
High-Pass Filter
Pressure Differential
Amplifier with
Protection Circuitry
Model Line Optical
Isolation for FECG,
IUP, US1 (174 only)
and TOCO
+4 V Reference
IUP Detection
Circuitry
Isolated Power
Supply
Filtering
and Gain
Stages
Gain and Offset
Adjustments
Filtering Stages
RA
LA
RL
PRESS
+PRESS
4 V
FECGEN
IUPEN
TOCOEN
Isolated
Supplies
Optical
Isolation
Optical
Isolation
To TOCO Enable
+12 V (173); +11.6 V (174)
FECGEN
IUPEN
TOCOEN
To Main Board Front End
FECG/IUP Connector
To Main Board A/D
Converter for Processing
Isolated Side
Connector
Unisolated Side
Connector
144 KHz (173 only)
Power Supply
Control PAL
144 KHz
CLEAR*
Q, Q*
174 Only
US1EN (174 only)
US1EN (174 only)

Table of Contents

Other manuals for GE Corometrics 170 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 170 Series and is the answer not in the manual?

GE Corometrics 170 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 170 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals