8-4 170 Series Monitor Revision C
2000947-004
Peripheral Devices: RS-232 Connectors
RS-232 Connectors
Two RS-232C serial interface receptacles (RJ-45) allow connecting the 170 Series
Monitor to the following devices:
Critikon Model 1846/1846SX Blood Pressure Monitor
Quantitative Sentinel/Perinatal System
Nellcor N-400 Fetal Pulse Oximeter
115-Compatible Device
Table 8-1 lists the factory default settings for each port. Refer to Customizing the
Monitor on page 4-10 for information on using the service setup mode to change
these settings.
Baud Rate
Select a baud rate that is compatible with the external device. The available settings:
600, 1200, 2400, 4800, 9600, 19,200, and 384,000 bps.
Mode
Select a communications mode compatible with the external device.
HP or HP with Notes
Select HP or HP w/notes for connecting to a Marquette Quantitative/Perinatal
System or another central system that uses the Hewlett Packard communications
protocol. HP mode does not print central station annotations on the fetal monitor
strip chart, whereas HP w/notes does.
Ext. BP
Select ext. BP for connecting to a Critikon Model 1846/1846SX Blood Pressure
Monitor.
Table 8-1. Communication Default Settings
RS-232 Port Intended Device Mode Baud Rate
1 QS System HP 1200
2 Critikon Model 1846/1846SX ext. BP 600

Table of Contents

Other manuals for GE Corometrics 170 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 170 Series and is the answer not in the manual?

GE Corometrics 170 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 170 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals