Revision C 170 Series Monitor 9-5
2000947-004
Maintenance: Preventative Maintenance Inspection
Preventative Maintenance Inspection
Equipment Required
Digital Multimeter
Dielectric Tester
Leakage Current Tester
Aluminum Foil
Ultrasound Test Body
(ultrasound transducer wrapped in aluminum foil)
Tocotransducer Test Body
(tocotransducer wrapped in aluminum foil)
FECG Test Body for Model 173 or 174 only
(Shorted legplate with all three leads tied together)
IUP Test Body for Model 173 or 174 only
(SensorTip transducer wrapped in aluminum foil)
Anti-Static Wristband
Static-Fee Work Surface
170 Series Operators Manual
Visual Inspection
Inspect the following for excess wear and/or signs of damage:
Power Supply Cord
AC-to-DC Converter
AC Line Cord
Ultrasound Transducer(s)
Legplate(s)
IUPC Transducer(s)
Remote Event Marker
Monitor Input Receptacles
Chassis
Membrane Switch Panel
Internal Harnesses/Connectors
Display Panel

Table of Contents

Other manuals for GE Corometrics 170 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 170 Series and is the answer not in the manual?

GE Corometrics 170 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 170 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals