Revision C 170 Series Monitor 6-25
2000947-004
Functional Checkout Procedure: Dual Heart Rate Test (Non-Pattern)
Table 6-6. Model 325 Simulator Settings for FECG/US Test
Section Switch Setting
FECG/MECG
Main RATE
Rate 120 BPM
Mode FECG
QRS Amplitude 50 µV
QRS Polarity +
ULTRASOUND/FMD
Mode US
Level MED
Rate RAMP
General Pattern Memory OFF