6-30 170 Series Monitor Revision C
2000947-004
Functional Checkout Procedure: Alarm Test
26. Use the simulator’s
Manual Adjustment knob to increase the US signal to 100
bpm. Verify the following on the monitor:
The alarm tone continues sounding.
The Alarm indicator stops continues flashing.
The affected heart rate numerics continue flashing.
27. Press the monitor’s front panel
Alarm Silence pushbutton. Verify that the
visual and audio alarm indications disappear.
28. This concludes the functional checkout procedure.