Revision C 170 Series Monitor 9-19
2000947-004
Maintenance: Preventative Maintenance Inspection
Dielectric (Hi-Pot) Tests
CAUTION
POWER OFF—Turn OFF the 170 Series Monitor prior to
performing any of the hi-pot tests.
Patient–to–AC-Line Using DC Voltage for One Minute
NOTE: The hi-pot tester voltage is 5.656 kVdc for these tests.
US
Attach an US test body to the monitor’s front panel US input. Connect from the foil to
line.
pass fail
US2
Attach an US test body to the monitor’s front panel US2 input. Connect from the foil
to line.
pass fail
TOCO
Attach a tocotransducer test body to the monitor’s UA input. Connect from foil to line.
pass fail
IUP
Attach an IUP test body to the monitor’s front panel UA input. Connect from foil to
line.
pass fail
FECG
Attach an FECG test body to the monitor’s front panel FECG input. Connect from the
leads to line.
pass fail