EasyManuals Logo

GE Corometrics 170 Series Service Manual

GE Corometrics 170 Series
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #221 background imageLoading...
Page #221 background image
Revision C 170 Series Monitor 9-19
2000947-004
Maintenance: Preventative Maintenance Inspection
Dielectric (Hi-Pot) Tests
CAUTION
POWER OFF—Turn OFF the 170 Series Monitor prior to
performing any of the hi-pot tests.
Patient–to–AC-Line Using DC Voltage for One Minute
NOTE: The hi-pot tester voltage is 5.656 kVdc for these tests.
US
Attach an US test body to the monitor’s front panel US input. Connect from the foil to
line.
pass fail
US2
Attach an US test body to the monitor’s front panel US2 input. Connect from the foil
to line.
pass fail
TOCO
Attach a tocotransducer test body to the monitor’s UA input. Connect from foil to line.
pass fail
IUP
Attach an IUP test body to the monitor’s front panel UA input. Connect from foil to
line.
pass fail
FECG
Attach an FECG test body to the monitor’s front panel FECG input. Connect from the
leads to line.
pass fail

Table of Contents

Other manuals for GE Corometrics 170 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 170 Series and is the answer not in the manual?

GE Corometrics 170 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 170 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals