13-8 Dash 2500 Patient Monitor 2042480-001 C
MASIMO SET® SpO2: Indications and Contraindications
WARNING
Remove nail polish and artificial nails. Placing a sensor on a
polished or an artificial nail may affect accuracy.
CAUTIONS
Patient safety:
Do not place any clip-on sensor in a patient’s mouth or on a
patient’s nose.
Do not place a clip-on finger sensor on a patient’s thumb or
across a childs foot or hand.
Observe the sensor site to assure adequate distal circulation.
Sensor sites should be checked as instructed in the MASIMO
Directions for Use supplied with the SpO
2
probe. Different sites
call for varying times between checks. Sensor sites should be
changed at least every 4 hours.
Parameter performance:
For best performance, place the sensor at heart level.
Placing a sensor distal to an arterial line may interfere with
adequate arterial pulsation and compromise the measurement
of SpO
2
.
Place the sensor so that the LEDs (light emitting diode) and the
photodiode are opposite each other.
4. Select SpO2 from Setup parameters in the Main Menu and set up the
desired monitoring parameters.

Table of Contents

Other manuals for GE Dash 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 2500 and is the answer not in the manual?

GE Dash 2500 Specifications

General IconGeneral
BrandGE
ModelDash 2500
CategoryMedical Equipment
LanguageEnglish

Related product manuals