2042480-001 C Dash 2500 Patient Monitor 14-11
Alaris® Turbo Temp®: Alarms
Alarms
Temperature Alarms
Name of Alarm Cause Message
TEMP Disconnected temperature probe becomes
disconnected
message appears in alarm message
field; parameter label flashes in vital
signs field; alarm forces value to dashes
TEMP Bad Probe the measured temp value is
determined to be out of limits
OR
bad probe
message appears in alarm message
field; parameter label flashes in vital
signs field; alarm forces value to dashes
TEMP Too Hot the measured temp value is
determined to be above the
upper limit
message appears in alarm message
field; parameter label flashes in vital
signs field; alarm forces value to dashes