Page 45 of 105 [EN] English – 109M4017 Revision -
Table 2-4
Table 2-5
Step Procedure (Vacuum)
Step Procedure (Pressure)
1 Set to vacuum operation
(-).
1 Set to Pressure operation
(+).
2 To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
2 To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster fully
counter clockwise.
3 Seal the system. 3 Seal the system.
4 Use the pump to set the
maximum vacuum or set
the vacuum you want to
adjust.
4 Use the pump to
pressurise the system (up
to ≈20 bar)
5 Adjust the vacuum
(+ decrease; - increase).
5 Open the refill valve (1
turn)
Increase the pressure
with the volume adjuster
If increasing pressure the
limit of travel is reached,
fully counter clockwise.
There is no change in
mechanism with the
clockwise until the
pressure starts to
Continue to do steps 7 to
10 until you get the