Page 61 of 105 [EN] English – 109M4017 Revision -
The difference between the two values is displayed as hysteresis
value for the switch. This utility can be used with Ramp Automation,
where the rising signal causes the switch to change state and the
falling signal causes the switch to resume its’ original state.
Figure 2-15 Switch Test Example
c. Relief Valve
This utility is only available with measure functions.
This utility tests circuits or mechanisms that have a cut-out response
when an input reaches a defined threshold value. The utility allows
the user to select a mode of operation which can be rising or falling.
The utility displays additional values that represent the maximum
and minimum values achieved by the input signal.