Illustration 6-44:
f. Connect the Fluoro UPS rectifier input cable from PDB CB5 as follows (Illustration 6-44):
Rectifier input phase 1 to X1-L1-1 terminal,
Rectifier input phase 2 to X1-L2-1 terminal,
Rectifier input phase 3 to X1-L3-1 terminal,
Rectifier ground to X1-PE-rectifier terminal.
g.
Connect the Fluoro UPS by-pass input cable from PDB CB7 as follows (
Illustration 6-44):
By-pass input phase 1 to X2-L1-2 terminal,
By-pass input phase 2 to X2-L2-2 terminal,
By-pass input phase 3 to X2-L3-2 terminal
By-pass neutral input to X1-N-Bypass terminal.
By-pass ground to X1-PE-Bypass terminal.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
234 20 PIST0021 - US Fluoro UPS and PDB connection