Table 2-4: Omega V Table
Configuration
Minimum Room length in mm (inches)
without Head Extender
with Head Extender
Angio 1278 mm (50.3 in)
6036 (237.6) 6036 (237.6) + 150 (6)
Cardio and Neuro 1395 mm (54,9 in)
6036 (237.6) 6036 (237.6) + 150 (6)
Table 2-5: Tilting table
Configuration
Minimum Room length in mm (inches)
without Head Extender
with Head Extender
Angio 1278 mm (50.3 in)
6136 (241.6) 6136 (241.6) + 150 (6)
Neuro 1395 mm (54,9 in)
6136 (241.6) 6136 (241.6) + 150 (6)
Table 2-6: Tilting table Upgrade
Configuration
Minimum Room length in mm (inches)
without Head Extender
with Head Extender
Angio 1278 mm (50.3 in)
6136 (241.6) 6136 (241.6) + 150 (6)
Neuro 1395 mm (54,9 in)
6136 (241.6) 6136 (241.6) + 150 (6)
Neuro (old distance, no longer used) 1595 mm
(62.8 in)
6175 (243.1) 6175 (243.1) + 150 (6)