Patient Interface Module(s).
•
Table Side User Interfaces (TSUI)
○
ISmart Box
○
Table Side Status Control
○
Table Panning Handle
○
ITU
NOTE:
No IPX4 components shall be installed on a cart.
•
Innova Central touch screen
•
In-room 3D Mouse
•
Accessories:
○
Table Head Extender
WARNING
THE TABLE HEAD EXTENDER SHALL NOT BE USED WHILE THE
TABLE IS TILTED. FAILURE TO DO SO MAY RESULT IN SEVERE
PATIENT INJURY.
WARNING
THE TABLE HEAD EXTENDER IS NOT COMPATIBLE WITH THE
INNOVA IQ WIDE TABLE TOP.
○
Armboard
○
IV pole
○
Quick Strap
○
Head holder: Attach Head Positioning device onto the head end of the table with the
head cradle pointing to the end of the table. The clamp angle pressure bar should be
resting on top of the table and the blue stabilizing feet under the table. Tighten the
clamping systems with the four- pronged knob. Test to check that head holder is
securely attached to the table. Adjust the height and angle of the head cradle by
loosening the long handled knob. Once cradle is set at desired position tightened the
long handled top. Place the patient’s head in the head cradle. Use the two velcro©
straps to hold the forehead and chin in place. Additional adjustment maybe made to the
position of the head cradle by loosening and tightening the long handled knob if desired.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
126 2 Room Layouts