Chapter 10 - Quality Assurance Page 10-3
Service at Indicated Time Daily Weekly Monthly Annually Notes
Clean Probes •* * or before each use
Clean Probe Holders •
Inspect AC Mains Cable • Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean Monitor and Touch Panel •
Check Control Panel Movement • Mobile Unit Check Daily
Console Leakage Current Checks •
also after corrective
maintenance or as required
by your facilities QA
program
Peripheral Leakage Current Checks •
also after corrective
maintenance or as required
by your facilities QA
program
Surface Probe Leakage Current Checks •
also after corrective
maintenance or as required
by your facilities QA
program
Endocavity Probe Leakage
Current Checks
Twice
Annually
Transesphongeal Probe Leakage
Current Checks
As Prescribed
in probe manual
Surgical Probe Leakage
Current Checks
As Prescribed
in probe manual
Measurement Accuracy Checks •
also after corrective
maintenance or as required
by your facilities QA
program
Probe/Phantom Checks •
also after corrective
maintenance or as required
by your facilities QA
program
Table 10-2 Periodic Maintenance Schedule