EasyManua.ls Logo

GE LOGIQ 100 PRO

GE LOGIQ 100 PRO
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
D
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Chapter 10 - Quality Assurance Page 10-3
Service at Indicated Time Daily Weekly Monthly Annually Notes
Clean Probes •* * or before each use
Clean Probe Holders
Inspect AC Mains Cable Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Console
Clean Monitor and Touch Panel
Check Control Panel Movement Mobile Unit Check Daily
Console Leakage Current Checks
also after corrective
maintenance or as required
by your facilities QA
program
Peripheral Leakage Current Checks
also after corrective
maintenance or as required
by your facilities QA
program
Surface Probe Leakage Current Checks
also after corrective
maintenance or as required
by your facilities QA
program
Endocavity Probe Leakage
Current Checks
Twice
Annually
Transesphongeal Probe Leakage
Current Checks
As Prescribed
in probe manual
Surgical Probe Leakage
Current Checks
As Prescribed
in probe manual
Measurement Accuracy Checks
also after corrective
maintenance or as required
by your facilities QA
program
Probe/Phantom Checks
also after corrective
maintenance or as required
by your facilities QA
program
Table 10-2 Periodic Maintenance Schedule
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 174 of 197

Table of Contents

Related product manuals