IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Chapter 3 - Installation Page 3-23
3-6-4 Peripherals Connections and Installation
3.6.4.1 FootSwitch Connection
The footswitch which is the remote FREEZE device is connected to the rear panel of the system.
this extra FREEZE switch is provided to enhance flexibility to freeze images when the system is not
within reach of the user.
Figure 3-18 Foot switch connection to rear panel
Pin Assignment of Foot Switch Connector
Installation of the Foot switch:
1.) Switch ON the LOGIQ™ 100 PRO unit.
2.) Connect the footswitch cable to the Footswitch slot provided in the rear panel.
3.) Press foot switch to check its functionality.
4.) By pressing the foot switch pedal the image in the screen should freeze.
Pin No. Signal
1 Foot Switch
2 GND
3 GND
4 GND
5 GND
Figure 3-19 Foot Switch Connector (Round 5-pin connector)