IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Page 4-4 Section 4-1 - Overview
Section 4-3 Functional Check Procedures
To perform these tests, you will need a convex or a linear probe. These procedures should be per
formed during installation. These procedures are also used as basic checks to use when service of
the system is required.
Turn “ON” the LOGIQ™ 100 PRO system. The 2 LEDS, External Video and Freeze blink and go off.
The Monitor displays the following. Refer figure 4-3 for details.
1.) B-Mode Image appears
2.) Gray Scale Display
3.) Horizontal Scale Marker
4.) Vertical Scale Marker
5.) Hospital Name (If previously entered)
6.) Date
7.) Time
8.) Probe Type
9.) Focus Marker on vertical scale
10.) Depth (150mm)*
11.) Gain (G54) *
12.) Dynamic Range (D54) *
13.) Frequency **.
* Points 10-12 form a continuous code in the lower left hand corner of the monitor and they are
factory default values.
** Only for C36 Probe.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 79 of 197