IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Page 10-12 Section 10-6 - Electrical Safety Tests
Test the system, peripherals and probes for leakage current. Excessive leakage current can cause
injury or death in sensitive patients. High leakage current can also indicate degradation of insulation
and a potential for electrical failure. Do not use probes or equipment having excessive leakage
current.
To minimize the risk that a probe may shock someone the customer should:
Not use a probe that is cracked or damaged in any way
Check probe leakage current:
* once a year on surface probes
* twice a year on endocavitary probes
* whenever probe damage is suspected
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 183 of 197