Page 10-20 Section 10-6 - Electrical Safety Tests
7.) Power ON the LOGIQ™ 100 PRO unit.
8.) Depress the LIFT GROUND rocker switch and record the highest current reading.
9.) Follow the test conditions described in Table 10-18 for every probe.
10.) Keep a record of the results with other hand copies of Planned Maintenance data.