GE HEALTHCARE
D
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Page 10-22 Section 10-6 - Electrical Safety Tests
10.6.6.5 Data sheet for Probe Source Leakage Current
The test passes when all readings measure less than the values shown in Table 10-14 and
Table 10-15. Record all data on the Planned Maintenance Inspection Certificate.
.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
Probe Tested:
Unit Power
Tester Power Polarity
Switch
Tester GROUND or
NEUTRAL Switch Measurement
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED
Table 10-18 Typical Data Sheet For Probe Source Leakage Current
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 193 of 197