IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 3 LOGIQ E9 Setup 3 - 39
3-9-1 Setup the LOGIQ E9 for DICOM Communications for R2.x.x Software (some features
only apply to R3.x.x) (cont’d)
Table 3-8 Utility -> Connectivity -> Device screen
1.
Select MyComputer.
2.
Assign an AE Title to the LOGIQ E9. (AE stands for Application Entity. DICOM services use
this to identify the LOGIQ E9.) AE Title is case-sensitive. This title may contain the Computer
Name from the TCP/IP page, if desired.
NOTE: It is NOT recommended to use the factory default. This is not prohibited, but more than
one system with the same AE Title can cause confusion.
3.
Edit the Port Number if needed. “104” is typical. Save your changes and reboot the system.