IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 8 Replacement procedures 8 - 23
8-4-6-11 Restoring Patient Archive and System Configurations
For complete instructions, refer to the latest revision of the LOGIQ E9 Basic User Manual, Chapter 16.
1.) From the Touch Panel, select Utility -> System -> Backup/Restore.
2.) On the Restore screen, select as needed:
- Patient archive to restore the patient archive images.
- Report archive
- System configuration to restore all System Settings/User Defined Configuration (Presets), OR
- One or several system configuration items to restore parts of the System Settings and User
Defined Configuration (Presets), see: Figure 4-21 "Backup/Restore settings" on page 4-28.
3.) Select the appropriate Source Device.
4.) If restore is done from a backup on a removable media, insert the media in the drive.
5.) Select Restore Now.
Depending on the selection, one or two Restore confirmation windows are displayed:
6.) Ensure that the correct source is selected, and select OK.
The selected items are copied to the systems. If items from the System Configuration are restored
the system needs to be rebooted. The Reboot system window is displayed.
7.) Select OK to reboot the system, and verify the Customer Presets are working properly.
8-4-6-12 Verifying the Video and DVR settings
1.) Select Utility -> System -> Peripherals.
2.) Verify Video Settings Format as recorded in Table 4-13 "Record settings from Peripheral screen"
on page 4-29.
Correct the selection if needed.
3.) Verify DVR model as recorded in Table 4-13 "Record settings from Peripheral screen" on page 4-29.
Correct the selection if needed.
4.) Format a DVD, and backup Customer Presets to confirm proper DVD/DVR write functionality.
CAUTION
The Restore procedure will OVERWRITE the existing data on the local Hard Drive. Make sure to select
the correct source device.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals