GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 8 Replacement procedures 8 - 23
8-4-6-11 Restoring Patient Archive and System Configurations
For complete instructions, refer to the latest revision of the LOGIQ E9 Basic User Manual, Chapter 16.
1.) From the Touch Panel, select Utility -> System -> Backup/Restore.
2.) On the Restore screen, select as needed:
- Patient archive to restore the patient archive images.
- Report archive
- System configuration to restore all System Settings/User Defined Configuration (Presets), OR
- One or several system configuration items to restore parts of the System Settings and User
Defined Configuration (Presets), see: Figure 4-21 "Backup/Restore settings" on page 4-28.
3.) Select the appropriate Source Device.
4.) If restore is done from a backup on a removable media, insert the media in the drive.
5.) Select Restore Now.
Depending on the selection, one or two Restore confirmation windows are displayed:
6.) Ensure that the correct source is selected, and select OK.
The selected items are copied to the systems. If items from the System Configuration are restored
the system needs to be rebooted. The Reboot system window is displayed.
7.) Select OK to reboot the system, and verify the Customer Presets are working properly.
8-4-6-12 Verifying the Video and DVR settings
1.) Select Utility -> System -> Peripherals.
2.) Verify Video Settings Format as recorded in Table 4-13 "Record settings from Peripheral screen"
on page 4-29.
Correct the selection if needed.
3.) Verify DVR model as recorded in Table 4-13 "Record settings from Peripheral screen" on page 4-29.
Correct the selection if needed.
4.) Format a DVD, and backup Customer Presets to confirm proper DVD/DVR write functionality.
The Restore procedure will OVERWRITE the existing data on the local Hard Drive. Make sure to select
the correct source device.