IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
8 - 68 Section 8-5 - Replacing Covers and Bumpers
8-5-15-1 Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
8-5-15-2 Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Connect cables and Probes you removed earlier
2.) Power up the system to verify that it operates as intended.
3.) LCD Arm and LCD Monitor movement functions in all directions without the covers loosening.
8-5-15-3 Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all required checks and is ready for use.
Table 8-46 LCD Monitor V2 Arm Assembly Covers replacement Functional
Checks
See:
Section
Functional Check Debrief Script
4-2-2 Power ON/Boot Up
LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all
required checks and is ready for use.
4-2-3 Power shut down
4-3-14 Probe/Connectors Checks
4-3-18 Mechanical Function Checks
4-3-18-2 Brakes and Direction Lock Checks