8-6-12-4 Gel Warmer removal
1.) Remove the Bulkhead Cover.
2.) Disconnect the Gel Warmer cable from (P1) on the back of the bulkhead.
3.) The Gel Warmer cable is routed through the cable channel (1) and the lower frame support (2). Pull
the Gel Warmer cable through the lower frame cable support to free the cable.
4.) Remove the Options Holder (8-6-11 "Options Holder / Left or Right Support replacement" on page
8-162).
5.) Remove the two screws at the Gel Warmer base.
6.) Lift the Gel Warmer out of the Options Holder.
Figure 8-48 Gel Warmer Cable placement
Figure 8-49 Options Holder removed, with Gel Warmer in place

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals