IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 8 Replacement procedures 8 - 201
8-9-2 BEP replacement - R3.x and earlier (cont’d)
8-9-2-1 BEP removal - R3.x and earlier
Table 8-122 BEP removal - R3.x and earlier
Steps Corresponding Graphic
WARNING
WARNING
When you return the used BEP to your local parts organization, make sure you remove ALL
PATIENT DATA from the Hard Drive, given that the Hard Drive is still functional.
In some countries, you may be required to delete all software from the disk before returning the
BEP to the parts warehouse. Follow your local policies.
Before you dispose of the Hard Drive, make sure you remove ALL PATIENT DATA from the Hard
Drive, given that the Hard Drive is still functional.
In some countries, you may be required to delete all software from the disk before returning the
Hard Drive to the parts warehouse. Follow your local policies.
CAUTION
When the BEP door is open, the Extended Power Shutdown circuit board is exposed. When
working inside the BEP, remove the battery cable form J3 on the Extended Power Shutdown.
Otherwise, the 24V battery will be exposed. If you drop a tool or a screw on the Extended Power
Shutdown, you could short the battery. Permanent damage will be done if this happens.
NOTE: Some cables in the BEP have a locking latch. Make sure to slide the locking latch towards you
and pull the connector to the left to release.
1.
Disconnect all I/O cables and all cables at the top of the BEP.
2.
Loosen the thumb screw or screws at the top of the BEP cover that attach the cover to the BEP.
Tilt the top of the BEP cover away from the BEP.
3.
Disconnect:
• the battery cable from J3 on the EPS
(Extended Power Shutdown) Board.
• the
cables (
1
) connecting the EPS to the
BEP.
Place the BEP cover aside.
inside of BEP Cover, EPS and cables

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals