D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 8 Replacement procedures 8 - 263
8-10-4-4-3 4D MC Replacement for Lambda PS
.
Table 8-167 4D MC Removal for Lambda PS
Steps Corresponding Graphic
1.
Locate the 4D MC Cover on the back of
the Main PS.
Remove the three screws from the
4D MC Cover. DO NOT discard. Gently
slide the 4D MC Cover and the 4D MC out
of the PS.
2. Remove the two screws (circled) that
secure the 4D MC to the 4D MC Cover.
Discard. Note how the 4D MC USB Cable
is routed for installation.
3. Disconnect the USB Cable from J28 on
the BEP.