GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 3 LOGIQ E9 Setup 3 - 35
3-7-5-2 Installing a Software Option
1.) From the Touch Panel, select Utility -> Admin -> System Admin.
2.) Enter the new option key code in the SW Option Key section.
3.) Select the Add button.
Figure 3-20 System Admin screen, R1.x.x - Add button
Figure 3-21 System Admin screen, R2.x.x or later - Add button
INCORRECT PASSWORD ENTRY WILL RESULT IN LOSS OF SYSTEM OPTIONS.
IF PASSWORD IS INCORRECT, PLEASE CONTACT YOUR LOCAL GE SERVICE
REPRESENTATIVE OR THE ONLINE CENTER.