General Procedures and Functional Checks
4-18 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Archiving and loading presets
NOTE: Always save presets before any software reload. This ensures
the presets loaded after the software reload are as up–to–date
as possible.
All user presets except changes to Summary, Anatomy, and
Biometry pages, can be saved on an DVD-R disk (or USB
memory device) for reloading on the system.
NOTE: Presets should NOT be saved on the same USB memory device
(or DVD-R disk) as images. The Archive Menu lists the images
but does NOT list the presets stored on a USB memory device
(or DVD-R disk, SD Card).
Archiving Presets to a USB memory device (or DVD-R Disk, SD Card)
1. Connect the USB memory device to the system’s USB port,
or insert an empty (blank) DVD-R disk into the DVD-RW.
2. Enter Utility -> System -> Backup/Restore. The Backup/
Restore sheet will be shown on the LCD display.
Figure 4-12. Backup Sheet