EasyManua.ls Logo

GE LOGIQ V1

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Ultrasound system requirements
LOGIQ V2/LOGIQ V1 – Basic Service Manual 2-11
5610739-100 English Rev.10
Probes environmental requirements
Operation, storage and transport temperatures for probes
Probes should be operated, stored, or transported within the
parameters outlined below.
NOTE: Refer to Table 2-1 on page 2-3 to determine the needed
settlement time.
Time and manpower requirements
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6: Probe Environmental Requirements
Operational Storage Transport
Temperature 10° - 40° C
50° - 104° F
-5° - 50° C
23° - 122° F
-5° - 50° C
23° - 122° F
Humidity 30- 80%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
CAUTION
Check the room temperature before you use the probe.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.

Table of Contents

Other manuals for GE LOGIQ V1

Related product manuals