General procedures
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-5
5610739-100 English Rev.10
Power ON/Boot Up
Warnings
DANGER
ALWAYS CONNECT THE ULTRASOUND SYSTEM TO A
FIXED POWER SOCKET WHICH HAS THE PROTECTIVE
GROUNDING CONNECTOR.
DANGER
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS
DEFEATS THE SAFETY GROUND.
DANGER
ENSURE THAT THE POWER CORD AND PLUG ARE
INTACT AND THAT THE POWER PLUG IS THE PROPER
HOSPITAL-GRADE TYPE (WHERE REQUIRED).
CAUTION
Ultrasound system requires all covers.
Operate this Ultrasound system only when all board covers and
frame panels are securely in place. The covers are required for
safe operation, good Ultrasound system performance and
cooling purposes.
CAUTION
Use only power supply cords, cables and plugs provided by or
designated by GE.