Care and Maintenance
10-6 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Maintenance task schedule
How often should maintenance tasks be performed?
The Care and Maintenance task schedule (provided in
Table 10-1 on page 10-6) specifies how often your LOGIQ V2/
LOGIQ V1 should be serviced and outlines items requiring
special attention.
NOTE: It is the customer’s responsibility to ensure the LOGIQ V2/
LOGIQ V1 care and maintenance is performed as scheduled in
order to retain its high level of safety, dependability and
performance.
Your GE Service Representative has an in-depth knowledge of
your LOGIQ V2/LOGIQ V1 and can best provide competent,
efficient service. Contact GE for coverage information and/or
price for service.
The service procedures and recommended intervals shown in
the Care and Maintenance Task Schedule assumes that you
use your LOGIQ V2/LOGIQ V1 for an average patient load
(10-12 per day) and not use it as a primary mobile Ultrasound
system which is transported between diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load,
then it is recommended to increase the periodic maintenance
frequencies.
Table 10-1: Customer Care Schedule
Service at Indicated
Time Daily Weekly Monthly
Per
Facilities
QA
Program
Notes
Clean Probes •* * or before each use
Inspect AC Mains Cable Mobile Ultrasound
system: Check
Weekly
Inspect Cables and
Connectors
Clean Console