EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-29
5610739-100 English Rev.10
Color Mode Checks
Introduction
Color Flow screens are 2D or M Mode screens with colors
representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode
or a combination of these.
Preparations
• Connect one of the probes to the scanner’s left-most probe
connector.
• Turn ON the scanner.
The B Mode window is displayed (default mode).
Select Color Mode
1. From an optimized B Mode image, press CF.
2. Use the trackball to position the ROI frame over the area to
be examined.
3. Press Scan Area. The instruction Size should be
highlighted in the trackball status bar. Use the trackball to
adjust the dimension of the ROI.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals