EasyManua.ls Logo

GE LOGIQ V1

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-29
5610739-100 English Rev.10
Color Mode Checks
Introduction
Color Flow screens are 2D or M Mode screens with colors
representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode
or a combination of these.
Preparations
Connect one of the probes to the scanner’s left-most probe
connector.
Turn ON the scanner.
The B Mode window is displayed (default mode).
Select Color Mode
1. From an optimized B Mode image, press CF.
2. Use the trackball to position the ROI frame over the area to
be examined.
3. Press Scan Area. The instruction Size should be
highlighted in the trackball status bar. Use the trackball to
adjust the dimension of the ROI.

Table of Contents

Other manuals for GE LOGIQ V1

Related product manuals