Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-39
5610739-100 English Rev.10
Back End Processor checks
If all the previous tests have been passed successfully, the
Back End Processor is most likely OK.
If the system seems to be operating erratically, please refer
to ‘Diagnostics/Troubleshooting’ on page 8-1.
Operator Panel Test
The Operator Panel is tested when the LOGIQ V2/LOGIQ
V1 is powered up as part of the start-up scripts, run at every
start-up.
For more info, please refer to ‘Diagnostics/Troubleshooting’
on page 8-1.
Peripheral checks
Printer checks
The internal printer is controlled from the Print and Store keys
on the LOGIQ V2/LOGIQ V1’s Operator Panel.
The factory default is:
Print for the standard printer
Store for the screen capture to clipboard
Turn OFF Power to LOGIQ V2/LOGIQ V1
See ‘Power off’ on page 4-9 for more information.