EasyManuals Logo

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #161 background imageLoading...
Page #161 background image
Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-39
5610739-100 English Rev.10
Back End Processor checks
• If all the previous tests have been passed successfully, the
Back End Processor is most likely OK.
If the system seems to be operating erratically, please refer
to ‘Diagnostics/Troubleshooting’ on page 8-1.
Operator Panel Test
• The Operator Panel is tested when the LOGIQ V2/LOGIQ
V1 is powered up as part of the start-up scripts, run at every
start-up.
For more info, please refer to ‘Diagnostics/Troubleshooting’
on page 8-1.
Peripheral checks
Printer checks
The internal printer is controlled from the Print and Store keys
on the LOGIQ V2/LOGIQ V1’s Operator Panel.
The factory default is:
• Print for the standard printer
Store for the screen capture to clipboard
Turn OFF Power to LOGIQ V2/LOGIQ V1
See ‘Power off’ on page 4-9 for more information.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals