EasyManuals Logo

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #217 background imageLoading...
Page #217 background image
Loading the software
LOGIQ V2/LOGIQ V1 – Basic Service Manual 7-27
5610739-100 English Rev.10
Software Upgrade Procedure (For R1.1.x)
This section is to describe software upgrade procedure about
how to upgrade the software of LOGIQ V2/LOGIQ V1 from
R1.0.x to R1.1.x and software upgrade between R1.1.x.
NOTE: If you are going to upgrade software from R1.0.x to R1.1.x, be
sure you have completed BIOS setting before you start software
upgrade procedure.
1. Insert the Upgrade USB memory stick labeled “System &
Application Software” in the USB port of the system.
2. Properly turn off the system by momentarily pressing the
Power On/Off Switch. Then select Shutdown in the System
Exit Window.
NOTE: If the system does not shutdown normally, hold down the
Power On/Off Switch until the light turns off
3. Power on the system and it will detect the USB memory
stick automatically.
4. Select the option Load the complete disk or Load the
bootable C: partition only for software upgrade. Or select
Exit to exit the upgrade process.
Figure 7-15. Upgrade instruction

Table of Contents