Common Diagnostics
LOGIQ V2/LOGIQ V1 – Basic Service Manual 8-21
5610739-100 English Rev.10
Assessment Utility
Probe Assessment
1. Place the probe’s carrying case on a stable surface and
open the case.
2. Carefully remove the probe and unwrap the probe cord.
3. Put the probe in the probe holder.
4. Align the connector with the probe port and carefully push
into place.
5. Lock the probe latch upward.
6. Carefully position the probe cord so it is free to move and is
not resting on the floor.
7. Enter global service user interface as GE service.
CAUTION
DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.