Site Preparations
2-4 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Environmental specifications for Ultrasound system
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Table 2-2: System Environmental Requirements
Operational Storage Transport (<16hrs.)
Temperature 10° - 40°C
50° - 104°F
-5° - 50°C
23° - 122°F
-5° - 50°C
23° - 122°F
Humidity 30 - 80%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
CAUTION
The LOGIQ V2/LOGIQ V1 system and probe connector is not
waterproof. Do not expose the device to water or any kind of
liquid.