EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
Site Preparations
2-20 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
DICOM option setup requirements
To configure the Ultrasound system to work with other network
connections, the site’s network administrator must provide
information to complete the form “Worksheet for DICOM
Network Information”. Ensure that there are no spaces in any
field of the form.
See:
Entries must include:
A host name, local port number, AE Title, IP address and
Net Mask for the Ultrasound system.
The IP addresses for the default gateway and other routers
at the site for ROUTING INFORMATION.
The host name, IP address, port and AE Title for each
device the site wants connected to the Ultrasound system
for DICOM APPLICATION INFORMATION. A field for the
make (manufacturer) and the revision of the device, is also
included. This information may be useful for error solving.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals