Report Formats
Item Description
7 Page number of the report. The page number increments for each page of the rhythm
report.
For the digital rhythm report, the page number is shown as Page <x> of <y>.
For the printed rhythm report, the page number is shown as Page <x>.
8 The type of report format (rhythm report).
9 The product name.
10 The Filter of the ECG waveform (measured in Hz), indicated with ZPD (for High Pass
filter).
NOTE:
You can change the filter before or during the recording and/or printing of a
rhythm.
For the printed rhythm report, the printing stops and restarts at the newly selected
filter. A gap is shown on the printed rhythm report where the change occurred. Each
time the rhythm printing starts after a change in filter, a calibration pulse is added
for each lead that tells which filter on the printed rhythm.
For the digital rhythm report, the calibration pulse of the last filter selected during
recording is shown on the full report.
11 The Gain of the ECG waveform (measured in mm/mV).
NOTE:
You can change the gain before or during the recording and/or printing of a rhythm.
For the printed rhythm report, the printing stops and restarts at the newly selected
gain. A gap is shown on the printed rhythm report where the change occurred.
Each time the rhythm printing starts after a change in gain, a calibration pulse is
added for each lead that tells the gain on the printed rhythm.
For the digital rhythm report, the calibration pulse of the last gain selected during
recording is shown on the full report.
12 The Speed of the ECG waveform (measured in mm/s).
NOTE:
You can change the speed before or during the printing of a rhythm.
For the printed rhythm report, the printing stops and restarts at the newly selected
speed. A gap is shown on the printed rhythm report where the change occurred.
Each time the rhythm printing starts after a change in speed, a calibration pulse is
added for each lead that tells the speed on the is printed rhythm.
The digital rhythm report is recorded at the configured rhythm speed.
13 The Leads.
2088531-370-2 MAC VU360
Resting ECG Analysis System 139

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours

Related product manuals