Page #37 background image
Patient Information
If Then
Multiple remote orders are found on the MUSE
system
The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.
One remote order is found on the MUSE system The order fills in the Patient Information screen, if
there is no mismatch with the patient data on the
screen.
No remote orders are found on the MUSE system A message displays in the notification area that
no matching remote orders are found. The device
then searches for ADT data in the MUSE system.
If multiple ADT records are found on the MUSE
system - The matching records show on the
Acquisition screen, and a message displays in
the notification area that matching ADT data
is found. Select the ADT record to search the
order using that selected record's Patient ID.
If only one ADT record is found on the MUSE
system - The device searches for the Patient
ID order found in that ADT record.
If multiple orders are found on the system
- The orders show on the Orders tab in
a filtered list. Select the order you will to
attach to the patient test.
If only one order is found on the system -
The order fills in the Patient Information
screen, if there is no mismatch with the
patient data on the screen.
If no orders are found on the system - A
message displays in the notification area
that no matching remote orders are found.
If remote order query is failed on the
system - A message displays in the
notification area that remote order query
has failed.
If no ADT records are found on the MUSE
system - A message displays in the notification
area that no ADT data/remote orders found.
If remote ADT data query failed on the MUSE
system - A message opens in the notification
area that ADT query has failed.
Remote order query failed on the MUSE system A message displays in the notification area that
remote order query has failed.
The device then searches for ADT data in the MUSE system with Visit Number.
2088531-370-2 MAC VU360
Resting ECG Analysis System 37

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours