Patient Information
If Then
The selected order matches the patient
demographics in the Patient Information screen
The following warning message opens on the
screen:
Are you sure you want to attach this order to the
current test?
Select Yes to attach the order, No to cancel the
action.
The selected order does not match the patient
demographics in the Patient Information screen
The following warning message opens on the
screen:
Are you sure you want to attach this order to the
current test?
Select Yes to attach the order, No to cancel the
action.
Orders and then ADT Query Workflow
The device first searches for local orders with the Patient ID.
If Then
Multiple local orders are found on the device The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.
One local order is found on the device The order fills in the Patient Information screen, if
there is no mismatch with the patient data on the
screen.
No local orders are found on the device A message displays in the notification area that no
matching local orders are found.
The device first searches for local orders with the Visit Number.
If Then
Multiple local orders are found on the device The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.
One local order is found on the device The order fills in the Patient Information screen, if
there is no mismatch with the patient data on the
screen.
No local orders are found on the device A message displays in the notification area that no
matching local orders are found.
The device then searches for remote orders in the MUSE system with Patient ID.
36 MAC VU360
Resting ECG Analysis System 2088531-370-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours

Related product manuals