Patient Information
The device then searches for remote orders in the MUSE system with Patient ID.
If Then
Multiple remote orders are found on the MUSE
system
The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.
One remote order is found on the MUSE system The order fills in the Patient Information screen, if
there is no mismatch with the patient data on the
screen.
No remote orders are found on the MUSE system A message displays in the notification area that no
matching remote orders are found.
Remote order query failed on the MUSE system A message displays in the notification area that
remote order query has failed.
The device then searches for ADT data in the MUSE system with Visit Number.
If Then
Multiple ADT records are found on the MUSE
system
The matching records show on the Acquisition
screen, and a message displays in the notification
area that matching ADT data is found.
The maximum of five records display on the
Acquisition screen.
Select the ADT record to search the order using
that selected record's Patient ID.
Only one ADT record is found on the MUSE system System searches for the Patient ID order found in
that ADT record.
If multiple orders are found on the system -
The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.
If only one order is found on the system - The
order fills in the Patient Information screen, if
there is no mismatch with the patient data on
the screen.
If no orders are found on the system - A
message displays in the notification area that
no matching remote orders are found.
If remote order query is failed on the system
- A message displays in the notification area
that remote order query has failed.
No ADT records are found on the MUSE system A message displays in the notification area that no
ADT data found.
Remote ADT data query failed on the MUSE
system
A message opens in the notification area that ADT
query has failed.
When local or remote orders are found, and you try to attach the order:
2088531-370-2 MAC VU360
Resting ECG Analysis System 35

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours

Related product manuals