Automatic Balance (for earlier software versions) F-1
January 2006
Programming Automatic
Balance for a VGA Optica
Note: For Optica Analyzers with software version XXX or later, see
Automatic Cleaning and Balance Function on page 4-9.
Programming Automatic
Balance for a 4x40
Optica
Note: For Optica Analyzers with software version XXX or later, see
Automatic Cleaning and Balance Function on page 5-8.
Table F-1: Automatic Balance Options
Item Function Available Options
Interval
Set how often the automatic balance takes place.
Enter the interval in minutes
(60 minutes minimum).
Type
Set the type of balance
(see Sensor Balancing on page 3-10).
Auto, PACER
Status
Set whether the balance will take place. Enabled, Disabled
Track/Hold
Set whether the display, analog outputs, and alarms
will track the sensor condition during a cleaning
cycle, or hold the last process reading.
(Serial outputs always hold the process reading).
Track, Hold
Table F-2: Automatic Balance Options
Setting Description Available Options
Interval in Minutes Set how often the balance takes place.
Enter the interval in minutes
(60 minutes minimum).
Type
Set the type of balance (see Sensor
Balancing on page 3-10).
Auto, PACER
Enable Set whether the balance will take place. Enabled, Disabled
Analog Output
Tracking
Set whether display, analog outputs and
alarms will track the sensor condition
during a cleaning cycle, or hold the last
process reading. (Serial output will
always track the process reading.)
Track, Hold