Programming the VGA Optica 4-9
January 2006
Automatic Cleaning and 
Balance Function
Note: For Optica Analyzers with earlier versions of software, see 
Appendix F.
Optica Analyzers with version XXX software have been upgraded to 
include the ability to program the PACER self-cleaning and 
rebalancing cycle to run once per day at a preset time. This is referred 
to as a Clock Time Interval. When the unit is programmed in this 
manner, the front panel PACER softkey is disabled. The Elapsed Time 
Interval initiates the PACER at a preset time after the last PACER was 
run. The automatic balance cycle will always run upon power up of 
the analyzers.
To program the Automatic Cleaning & Balance Function on the 
Optica VGA monitor:
1. Make sure the internal clock of the Optica has the correct time.
a. From the main screen enter Menu, then More, then Set Time 
and Date.
b. Highlight the Date and Time fields and use the Keypad and 
Softkeys to set the correct time.
2. Set the Time Programmed Balance Cycle.
a. Enter Menu and go to the Automatic Balance Section.
b. Open the Type pull down menu and choose one of the four 
selections (see Table 4-6 below).
Note: The PACER function will provide more thorough cleaning than 
the AUTO function because it first develops a thick dew/frost 
layer into which soluble contaminants dissolve. When heated, 
some of the contaminants are flash-evaporated and the 
remaining residue accumulates in clusters, resulting in an 
approximately 85% cleaner surface. The PACER cycle 
generally takes longer to complete.
Table 4-6:  Automatic Cleaning and Balance Functions
Item Function Type
AUTO Heats the mirror and balances the optics. Elapsed Time Interval
PACER
First cools the mirror to develop a thick dew/frost 
layer, then heats and balances the optics.
Elapsed Time Interval
AUTO-D Heats the mirror and balances the optics. Clock Time Interval
PACER-D
First cools the mirror to develop a thick dew/frost 
layer, then heats and balances the optics.
Clock Time Interval